J457 Refractometer

J457 Refractometer


 J457 AUTOMATIC REFRACTOMETER

Visit the Rudolph Research Analytical website for additional information

Download the bochure (2MB).pdf

Smart Measure™ Technology Improves Results!

All J457's come with Smart Measure™ Technology that automatically detects when the prism is improperly cleaned, insufficient sample is loaded, or if the instrument is improperly calibrated.

Exclusive Dual Temperature Control System

All J457 models come with Rudolph's exclusive Dual Temperature Control System utilizing Peltier Technology. The J457 heats and cools the sample from both prism surface and sample cover. The insulating cover housing seals the Prism Well Ring to create a uniform temperature environment providing unmatched temperature control uniformity.

The J457 comes in three configurations to match your operating conditions

J457-SC Standard Configuration

 

J457-SC Standard Configuration is Rudolph's most popular configuration and provides an excellent combination of a small footprint and a large, easy-to-navigate touchscreen interface.

 

J457-WC Wall Mount Configuration

J457-WC Wall Mount Configuration is designed for production environments where water is used for cleaning or there is a lot of sample spillage on the work surface. This configuration is also excellent for situations where there is very little work space available: out on the factory floor or in an over crowded satellite laboratory where there is little to no remaining bench space.

 

J457-FC Factory Configuration

J457-FC Factory Configuration is recommended when there is a lot of sample spillage on the laboratory's instrumentation. Factories working with sticky resins and syrups that end up being spilled on the instrumentation are great candidates for the J457-FC as the measurement unit can be as much as six feet away from the display unit.

 

The J457 Smart Measure™ Advantage

Measurement problem:
> Sample residue is left on prism surface

 
Experience has shown that incorrect measurements are often linked to cross contamination caused by the previous sample being inadvertently mixed with the current sample being measured. Ensuring that the prism surface is properly cleaned between measurements can be difficult for busy environments where multiple operators are involved.

Smart Measure™ solution:
> Automated clean prism monitoring

Smart Measure™ knows when the prism is clean. If the prism surface does not produce the correct "Clean Prism" result then Smart Measure™ notifies the operator that further cleaning of the prism is necessary. Smart Measure™ notifies the operator that it is ready to measure after the prism is satisfactorily cleaned.

Measurement problem:
> An insufficient amount of sample covering prism surface

 
A small drop of sample in the center of the prism will measure correctly. A large drop of sample completely covering the prism surface will measure correctly. A small drop on the perimeter of the prism will produce erroneous results.

Smart Measure™ solution:
> Automated sample load monitoring

Smart Measure™ knows when a sample has been loaded on the prism and immediately begins evaluating various metrics surrounding the sample load. If Smart Measure™ determines that those metrics indicate proper sample loading, it will display the result with a sample quality number. If Smart Measure™ determines that the metrics indicate an improper sample load, Smart Measure™ will ask the operator to clean the prism and re-load the sample.

Measurement problem:
> Sample contamination or improper calibration

 
An improperly cleaned prism, where sample residue is left in the measurement well, can mix with a water zero or calibration fluid. This type of contamination during the calibration process will cause an error across all subsequent measurements.

Smart Measure™ solution:
> Automated sample calibration inspection

Smart Measure™ is constantly measuring the prism surface and monitoring the value measured against its database of results. If during a water zero the instrument detects a value that is too different from the database or the RI value being input is far from the measured value, it will alert the operator.

Measurement problem:
> The sample has air bubbles that affect measurement results

 
Many samples have air trapped inside them or are carbonated. These samples read well most of the time as long as air bubbles are not trapped between the prism surface and the sample. When this happens the operator gets an erroneous result often referred to as a flyer. These results are called flyers because the measured value is quite different from the expected measurement value.

Smart Measure™ solution:
> Automated sample calibration inspection

Smart Measure™ catches these flyers by scanning the detector for varying RI values and then alerts the user to reload the sample. Since the user does not get a value to record until a new sample load is completed, Smart Measure™ greatly reduces the erroneous measurements from being recorded.

Smart Measure™ Features
Smart Measure™ knows when to measure and display a result even when the operator does not.

Smart Measure™ Features J457 (all models)

  • Dirty prism detection
  • Incorrect calibration detection
  • Insufficient sample load detection
  • Cross contamination detection
  • Improper water zero detection
  • Trapped air bubble on prism surface detection
 
Yes
 
No
 
Yes
 
No
 
Yes

J457 Automatic Refractometer Specifications (all models)

 

 
Measurement Scales Refractive Index (nD)
BRIX (% sucrose)
and up to 100 custom programmed scales
Measurement Range Refractive Index 1.26 - 1.70,
BRIX 0 - 100
Accuracy Refractive Index ±0.00002,
BRIX ±0.15
Reproducibility Refractive Index ±0.00002,
BRIX ±0.015
Resolution Refractive Index 0.00001
BRIX 0.01
Temp. Control Range 10°C to 100°C (wider temperature range available dependent on ambient conditions and options).
Temperature controlled by dual Peltier system.
Temp. Control Reproducibility ±0.01°C
Ambient Temp. Limit 5°C to 40°C
Temp. Correction Range 4°C to 95°C (for sucrose applications)
Sample Temp. Limit -20°C to 250°C
Optical Wavelength 589.3nm (NaD line)
Response Time User configurable, generally 30 seconds
Calibration Using water or NIST traceable fluids. Factory default calibration can always be reset
Prism Artificial Sapphire
Acid Resistance Hastelloy™ measurement surface (optional)
Data Storage/Internal Memory 8GB non-removable Compact Flash
Display 10.4 inch diagonal, 800x600 pixels, color, Flat Panel Monitor with Resistant Touch Screen Interface, 400 nits brightness, gasketted for spill protection.
User Interface Touchscreen
Communication Interface 3 USB, RS232 and Cat5 Network (Ethernet)
Operating Dimensions/Weight L: 17 ¼”, W: 12”, H: 13” / 23 lbs.
L: 43.5 cm, 30.5 cm, 33 cm / 10.4 kg
Power Requirements 100 - 240 volts, 50 Hz - 60 Hz

Alternative instruments

J57
J157
J257
J357

You may also be interest in

ECS  (Easy Clean Sampling System)

 

Special Offer

We currently have demo equipment available for the Refractometers, Density Meter, Smartchem and SPE Autosampler.

What's New

18/11/2008

  DDM 2911 Density Meter is available for demonstration

We are pleased to announce that Steroglass STRIKE 300 Rotary Evaporator won the GOLD AWARD as the most innovative and well designed instrument in the world according the IBO magazine. They declare that "The Strike 300's innovative industrial design clearly distinguishes it by creating an appealing visual appearance and enhanced end-user experience."

 

Phone: 02 9543 7377
Mobile: 0418 489 842
Fax: 02 9543 7366
Email: sales@alpe.net.au

Service NSW 0402 473 740
Service QLD 07 5465 6306
Service SA    0419 800 918
Service Vic     0408 305 325